HORIZONTAL LAMINAR AIR FLOW
Code: 25011-00
Description: LAMINAR AIR FLOW - HORIZONTAL MODEL ATH 1500/ ATH-1200
*Over all dimensions ( ATH 1500) : 1510 x 610 x 610 mm
( ATH-1200) : 1210 x 610 x 610 mm
*Direction of Flow Horizontal
*Particle Retention 0.3 Micron & Above
*Cleanliness level CLASS 100 all along the working area
Size: 4’ x 2’x 2’, Cleanliness Class 100.
*Sound Emission / Noise Level <60 dBA according to EN 12469
and NSF Standards.
*Fluorescent light intensity at zero ambient >1400 Lux
*Cabinet Construction
Main Body : will be 1.2 mm, 16 Gauge SS with Matt finish.
Electrical : 220 – 240 V AC 50 Hz, single phase
*Standard Compliance
Equipment quoted is as per ISI Guidelines / as per US FED 209 E STD
Type tested to Indian standards as equivalent to EN 12469:2000
*For Air Quality / Air Cleanliness
ISO 14644-1 CLASS 3 standards and other equivalent air cleanliness
requirements
*HEPA Filter – Minipleat Type Construction
Media : Ultra clean filter fiber paper – imported
Casing : Epoxy coated Aluminum frame
Retention : 0.3 Micron
Efficiency : 99.97%
Pressure Drop : 12 mm
*PRE Filter:
Media : Synthetic, non-woven polyester fibers, HDPE - Washable
Casing : Enamel painted CRCA frame
Separators : Aluminum separators
Retention : 10 Micron
Pressure Drop : 6mm WG
Grade F7 rating
* Blower Assembly:
Supply air blower: By EBM German imported Blower statically
and dynamically balanced centrifugal impeller. Noise level will
be 55 dBA. Four speed regulator will be provided.
*The sides will be covered by clear glass / acrylic.
*Front door will be of acrylic with double foldable type.
*The following tests will be undertaken at the time of installation and
corresponding calibration certificate will be provided.
Particle Count Test
Light Intensity test
Noise Level Test
* Do * Same as above but With Wooden Construction.