DUAL CHANNEL FLAME PHOTOMETER (2F) 392
Code: 08044-05
Description: Digital Clinical Flame Photometer Model 391 is a DualChannel Flame Photometer used for simultaneous observation of Na & K in blood serum (Plasma), Urine & other type of biological fluids.
TECHNICAL DATA
Serum Urine
RANGE Na : 0-200 meq/I 0-200 meq/ I
K : 0-10 meq/I 0-100 meq/I (Optional)
Ca : - 0-10 meq/I
Li : 0-2 meq/I (Optional)
READOUT : Dual 2 / digit bright red LED display
ACCURACY : ±2% upto 100 meq,±5% above 100 meq
IGNITION SYSTEM : In-built electronic Ignition by press of a switch
DIMENSIONS : 375 x 245 x 220 mm (Approx.)
MEASUREMENT SYSTEM : 2 elements on dual LED display
REPEATABILITY : ± 2 Counts
DETECTOR :Silicon Photodiode
FILTERS : Narrow band interference glass filters
NEBULIZER : Black bakelite, axial flow type
FLAME SYSTEM : LPG & dry oil free air
WARM UP TIME 10 minutes
POWER : 230 V±10% AC, 50 Hz
WEIGHT :Main Unit : 7.5 Kg. (Approx.) Compressor : 8 Kg. (Approx.)
COMPRESSOR UNIT
Air Supply : By oil free mini compressor
unit with pressure regulator
Combustion Gas LPG : Controlled by
precision regulator
Power : 230 V ± 10% AC, 50 Hz
Dimensions (LxBxH) : 290 x 255 x 210
mm (Approx.)
Weight : 8 Kg. (Approx.)