COD THERMO REACTOR RD125

COD THERMO REACTOR RD125

Code: 12016

Description:

COD THERMO REACTOR RD125 

The chemical oxygen demand, ST-COD value (small scale sealed tube), of water as determined by this dichromate method, can be considered as an estimate of the theoretical oxygen demand, i.e. the amount of oxygen consumed in total chemical oxidation of the organic constituents present in the water.

The Lovibond COD VARIO setups allow highly sensitive and precise water testing with minimum effort. They measures the ST-COD concentration by photometric detection employing a linear relationship between absorbance and concentration. After adding the sample to a Lovibond COD VARIO tube test (LR, MR, HR according to ISO 15705:2002), it is heated in the reactor and then analysed in the photometer.

SPECIFICATION

* Test Range : COD LR : 0-150mg/l

COD MR: 0-1500mg/l

COD HR : 0-15000mg/l

* ABS housing, PPS Protection grid, PC Lid, PBT Block insert

* Heating block made of Aluminium

* 24 Holes in the aluminium block of 16.2 mm ±0.2 mm

* Selectable temp. 100 / 120 / 150 ーC

* Probe type Pt100 A class

* Temperature stability ± 1C at the Pt 100.

* Selected time 30 / 60 / 120 / min. and continuous operation

* Heating up from 20C to 150C in 12 min.

* Regulation Microprocessor

* Protection against overheating at the aluminium block at 190C

* Environmental conditions 10 – 40C, max. 85 % rel. humidity

* Power supply 230 V / 50-60 Hz or 115 V / 50-60 Hz (switchable)

* Power 550 W

* Dimensions 248 x 219 x 171 mm

* Weight 3.9 kg

* CE-Conformity